Abstract:

In the bio-mining industry there are various compounds which need to be frequently monitored, one of which is iron, since its concentration can seriously affect the industrial process. This study describes the development of an electrochemical method of speciation and rapid quantification of iron in bio-mining processes conditions. Speciation and quantification was carried out in a flow injection system using techniques of cyclic voltammetry and chronoamperometry. The proposed and a standard method were used to compare the results obtained from simulated and real mixed (ferric/ferrous) samples. Cyclic voltammograms presented two peaks corresponding to the ferric ion reduction of and ferrous ion oxidation. Chronoamperometry allow the construction of calibration curves for both ions. Quantification of real and simulated samples showed acceptable levels of error compared to a standard method. The electrochemical method allowed in a simple and rapid way the speciation and quantification of iron in mixed samples, including real samples from acid mine drainage, which established the basis for the development of an automated process monitoring device. © 2014 IEEE.

Registro:

Documento: Conferencia
Título:Flow-injection amperometric sensor for quantification and speciation of iron
Autor:Saavedra, A.; Cortón, E.; Garcia L.
Filiación:Laboratory of Biosensors and Bioanalysis (LABB), Departamento de Química Biológica and IQUIBICEN-CONICET, Universidad de Buenos Aires, Argentina
Palabras clave:Electrochemistry; iron; speciation and quantification; Amperometric sensors; Automation; Chronoamperometry; Cyclic voltammetry; Electrochemistry; Ions; Iron; Process monitoring; Acid mine drainage; Calibration curves; Cyclic voltammograms; ELectrochemical methods; Ferrous ion oxidation; Flow-injection system; Industrial processs; speciation and quantification; Iron compounds
Año:2014
DOI: http://dx.doi.org/10.1109/IBERSENSOR.2014.6995519
Título revista:2014 IEEE 9th IberoAmerican Congress on Sensors, IBERSENSOR 2014
Título revista abreviado:IberoAmerican Congr. Sensors, IBERSENSOR - Conf. Proc.
Registro:https://bibliotecadigital.exactas.uba.ar/collection/paper/document/paper_97814799_v_n_p_Saavedra

Referencias:

  • Acevedo, F., Gentina, J., Process engineering aspects of the bioleaching of copper ores (1989) Bioprocess Enginnering, 4, pp. 223-229
  • Bastias, M., Gentina, J., Variables affecting the growth and ferrous oxidation capacity of L. Ferrooxidans in continuous culture (2010) Hydrometallurgy, 104, pp. 351-355
  • Muir, M.K., Anderson, T., Determination of ferrous iron in copper process metallurgical solutions by o-phenanthroline colorimetric methods (1977) Metallurgical and Materials Transactions B, 8, pp. 517-518
  • Huang, S., Zhou, L., Fe2+ oxidation rate drastically affect the formation and phase of secondary iron hydroxysulfate mineral occurred in acid mine drainage (2012) Materials Science and Engineering C, 32, pp. 916-921
  • Córdoba, E.M., Muñoz, J.A., Blázquez, M.L., González, F., Ballester, A., Leaching of chalcopyrite with ferric ion. Part I: General aspects (2008) Hydrometallurgy, 93, pp. 81-87
  • Silverman, M.P., Lundgren, D.G., Studies on the chemoautotrophic iron bacterium Ferrobacillus ferrooxidans: An improved medium and a harvesting procedure for securing high cell yields (1959) Journal of Bacteriology, 77, pp. 642-647
  • Druschela, G.K., Bakera, B.J., Gihring, T.M., Banfield, J.F., Acid mine drainage biogeochemistry at Iron Mountain, California (2004) Geochemical Journal, 5, pp. 13-32
  • Min, L., Rees, N.V., Kabakaev, A.S., Compton, R.G., Determination of iron: Electrochemical methods (2012) Electroanalysis, 24, pp. 1693-1702
  • Olsson, L., Schulze, U., Nielsen, J., On-line bioprocess monitoring-an academic discipline or an industrial tool? (1998) Trends in Analytical Chemistry, 17, pp. 88-95
  • Cranny, A., Harris, N.R., Nie, M., Wharton, J.A., Wood, R.J.K., Stokes, K.R., Sensors for corrosion detection: Measurement of copper ions in 3.5% sodium chloride using screen-printed platinum electrodes (2012) IEEE Xplore: Sensors Journal, 12, pp. 2091-2099A4 -

Citas:

---------- APA ----------
Saavedra, A., Cortón, E. & Garcia L. (2014) . Flow-injection amperometric sensor for quantification and speciation of iron. 2014 IEEE 9th IberoAmerican Congress on Sensors, IBERSENSOR 2014.
http://dx.doi.org/10.1109/IBERSENSOR.2014.6995519
---------- CHICAGO ----------
Saavedra, A., Cortón, E., Garcia L. "Flow-injection amperometric sensor for quantification and speciation of iron" . 2014 IEEE 9th IberoAmerican Congress on Sensors, IBERSENSOR 2014 (2014).
http://dx.doi.org/10.1109/IBERSENSOR.2014.6995519
---------- MLA ----------
Saavedra, A., Cortón, E., Garcia L. "Flow-injection amperometric sensor for quantification and speciation of iron" . 2014 IEEE 9th IberoAmerican Congress on Sensors, IBERSENSOR 2014, 2014.
http://dx.doi.org/10.1109/IBERSENSOR.2014.6995519
---------- VANCOUVER ----------
Saavedra, A., Cortón, E., Garcia L. Flow-injection amperometric sensor for quantification and speciation of iron. IberoAmerican Congr. Sensors, IBERSENSOR - Conf. Proc. 2014.
http://dx.doi.org/10.1109/IBERSENSOR.2014.6995519